top of page

Rapidox 6100 QU4D multi-gas pump back analyser

Exceptional accuracy and stability are provided when measuring the purity of the compartment gas, through specially selected sensors. The modular configuration allows for up to eight compatible gases to be analysed in each gas type, simultaneously, using one analyser.


The Rapidox is fully compatible with mixtures of SF6 & CF4, C4-FN & CO2/O2, C5-FK & CO2 and Clean Air mixtures of N2 and O2, together with toxic contamination gases such as SO2, HF, H2S & CO in SF6, CO C4-FN & C5-FK and NO2 in Clean Air. The unit also measures the water content of the gas in dewpoint or ppm to ensure dryness is acceptable.


The Rapidox 6100 QU4D is housed neatly into a tough Peli transport case and supplied with sample hoses, DN8 and DN20 coupling accessories. Once powered and connected, the operator can select which gas type to work with and the Rapidox configures the measurement screen correctly. If the user is changing gas type, the unit will run a purge cycle to clean the existing gas from the analyser and into an external recovery bag. During the measurement, the Rapidox automatically removes a small quantity of pressurised gas from the electrical equipment or gas cylinder, controlled with an auto gas pressure sensing function.


A vacuum purge cycle and internal gas storage system ensures that no air can contaminate the gas sample and that no test gas is able to escape during the testing period. Safety features are built in to ensure the cycle is completed correctly without gas loss or cross contamination. The analyser is pre-programmed with all current IEC and CIGRE test configurations, and has the ability to create customised test parameters.

  • Technical Specifications


    Specifications

    Gas Sample

    SF6 | C4-FN | C5-FK | Clean Air

    Measurement Time

    2 -8 minutes (Min 6 minutes required for H20)

    Gas Flow Rate

    0.5L.min-1

    Pressure Range

    0.3 - 35 Bar; displayed on screen

    Max Inlet Pressure

    35 Bar gauge

    Display

    7” full-color LCD touch screen with soft menu keys

    Sensor

    SF6, Air/N2,H2O,SO2,HF,CF4,H2S,CO,CO2,C4-FN&C5-FK and NO2

    Customizable sensor options available

    Coupling

    RAGON DN08 & DN20 with Rectus style fitting

    Printer

    Integrated thermal printer

    Ambient Operating Conditions

    -10°C to +40°C, 10-90%RH

    Data Storage

    4 GB internal data storage

    Warm-up Time

    Min 3-4 minutes at 20 deg. C

    Battery Life

    Up to 8 hours. 2–4-hour charge

    Power Voltage (Charging)

    90-260 VAC, 50/60 Hz

    Pump Back

    With

    Analyser Dimensions

    270mm(H) x 560mm(W) x 450mm(D)

    Weight

    21kg (Total instrument and case)


bottom of page